Siemens AG SPR2-310.815.01.07.02 ARCADIS Varic
12.06 CS PS SP
Final Work Steps 49
Page 49 of 64
Medical Solutions
Mark the following rows in the "Device leakage current" protocol, column "First mea-
sured value", with "n.a." as "not applicable":
¹ Row "Meas. inst. calibrated up to:"
¹ Row "Evaluation:"
Complete the following rows in the "Device leakage current" protocol, column "First
measured value":
¹ In the row "Date:" enter the current date.
¹ In the "Name" row, enter your name in block letters.
¹ Sign the protocol in the "Signature" row.
If you want to complete the protocol with additional remarks, enter them in the
"Remarks" table in the protocol addendum.
Device leakage current measurement during start-up (if necessary)
NOTE
If the programmed line voltage has been changed from its status
at delivery, and/or there have been changes of parts in the sys-
tem's primary circuit, the device leakage current value specified
in test protocol 1 delivered with the system is no longer valid.
The device leakage current measurement has to be performed.
NOTE
Observe the leakage current measurement instructions in the
"General Information" chapter of these instructions.
WARNING
Electrical voltage!
Non-compliance can lead to severe injury and even death.
¹ The system leakage current measurement may be per-
formed on systems of protection class I only after the
protective conductor test has been passed.
Perform the device leakage current measurement after successfully completing the
protective conductor test.
Complete the "First measured value" column of the "Device leakage current" protocol.
If you want to complete the protocol with additional remarks, enter them in the
"Remarks" table in the protocol addendum.

Table of Contents

Other manuals for Siemens ARCADIS Varic

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens ARCADIS Varic and is the answer not in the manual?

Siemens ARCADIS Varic Specifications

General IconGeneral
TypeMobile C-arm
Image Intensifier9-inch or 12-inch
DICOM CompatibilityYes
Frame RateUp to 30 fps
Focal Spot Size0.3/0.6 mm
MonitorDual 18-inch TFT LCD monitors

Related product manuals