Compact Operating Instructions in accordance with EN 61010-1 and EN 60079-0
20 A5E45779144002-01, 09/2018
Toxic and corrosive gases
Appropriate components must be used for toxic and/or flammable gases. These components must be approved for the
respective application.
We recommend that you consult our technical department before connecting toxic or corrosive gases.
Please note in particular that the control cabinet housing the analyzer must have a protection rating of at least IP54 and be
purged with a suitable purging gas.
Connection diagrams
ULTRAMAT 23
-X45: ELAN (RS485) 9-pin connector
-X90: 9-pin interface connector (option board with
-X50: 37-pin connector: Option board; binary in-
-X80: 37-pin connector: Analog and digital inputs and
When installing in a cabinet, mount analyzer on support rails
Figure 4-7 ULTRAMAT 23, gas connections and electrical connections
Figure 4-8 ULTRAMAT 23, assignment of gas connections