Compact Operating Instructions in accordance with EN 61010-1 and EN 60079-0
A5E45779144002-01, 09/2018
33
7.8
Technical specifications FIDAMAT 6
Enclosure
Weight
Approx. 23 kg
Dimensions (W x D x H)
483 mm x 457 mm x 160 mm
Electrical characteristics
Power consumption Approx. 150 VA during operation
Approx. 350 VA in the warm-up phase
Fuse values
100 to 120 V
200 to 240 V
4.0 T/250
2.5 T/250
Gas inlet conditions
Sample gas pressure
With integrated pump
Without pump
600 ... 1100 hPa
< 2000 hPa
Sample gas temperature
0 ... 200 °C
Sample gas humidity
< 90% relative humidity
Pressures and flow (FIDAMAT 6 with pump)
Gases
Inlet pressure hPa
(rel.)
Operating pressure
hPa (rel.)
Flow across FID
ml/min
Flow through by-
pass
ml/min
Without pump
start-up
With pump start-up
Combustion gas
2000 ... 4000
1000 ± 20
~ 12
-
Combustion air
2000 ... 4000
420 ± 20
500
~ 320
~ 500
Sample gas
atmospheric
-
500 ± 2
~ 3
~ 1000
Zero gas
2500 ... 3000
-
1500 ± 2
~ 3
~ 1000
Calibration gas
2500 ... 3000
-
1500 ± 2
~ 3
~ 1000
Pressures and flow (FIDAMAT 6 without pump)
Gases
Inlet pressure
hPa (rel.)
Operating pressure
hPa (rel.)
Flow across FID
ml/min
Flow through by-
pass
ml/min
Without sample
gas, calibration
gas
With sample gas, cali-
bration gas
Combustion gas
3000 ... 5000
2000 ± 20
~ 25
-
Combustion air
3000 ... 5000
1480 ± 5
--
~ 320
~ 300
Sample gas
1500 ... 2000
-
1500 ± 2
~ 3
~ 500
Zero gas
1500 ... 2000
-
1500 ± 2
~ 3
~ 500
Calibration gas
1500 ... 2000
-
1500 ± 2
~ 3
~ 500