Hematek
®
3000 System – Instruction Manual – Version 1.326
6 Cleaning
6
6.1 Periodic cleaning
Periodic cleaning actions are listed in Table 6-1.
6.2 Cleaning the exterior surfaces
To clean the exterior surfaces of the system, proceed as follows:
1. Wipe the exterior surfaces of the system with a damp, soft, disposable,
lint-free, absorbent cloth or tissue and mild cleaning agent.
2. Use a small amount of methanol to clean stain from the exterior
surfaces.
6.3 Cleaning the stain tube and cannula
To clean the stain tube and cannula, proceed as follows:
Remove cannulas 1. Remove the stain, buffer, and rinse cannulas from the Hematek Stain
Pak bottles.
Interval Action See
Daily, end of shift, or if necessary Clean exterior surfaces Chapter 6.2
Daily, end of shift Clean stain tube and cannula Chapter 6.3
Daily, end of shift, or if no glass slides are to be
processed for one hour or more
Clean platen Chapter 6.4
Daily, end of shift, or after installing a new
Hematek Stain Pak
Clean waste tank Chapter 6.5
Weekly Clean drain troughs and rear guide
rail
Chapter 6.6
Table 6-1 Periodic cleaning list
CAUTION
Cleaning agents can be harmful or irritating to your skin, eyes, etc.
Wear appropriate personal protective equipment; that is, gloves,
protective clothing, safety glasses and mask.
Pay attention to the manufacturer's information regarding the gloves'
compatibility with the materials and liquids used.
Follow all product manufacturer’s warnings.
Note: The buffer and rinse cannulas and tubes do not require cleaning.
These cannulas are removed so the buffer and rinse are not wasted
while the stain tube and cannula are being cleaned.