2. Check all connections to make sure the underplaten tubing is securely
in place.
3. Set the waste tank back in place.
8.5.3.2 Recorded volumes do not fall within specified ranges
If the rinse volume deviates significantly from 10mL for 10 blank glass slides
and/or the stain-to-buffer ratio is not between 1:2 and 1:3, proceed as
follows:
1. If you already corrected the pump volumes, contact the Siemens
service.
2. Otherwise, connect the tubes to the spouts under the platen. The tube
from the stain pump must lead to the right spout. The tube from the
buffer pump must lead to the middle spout. The tube from the rinse
pump must lead to the left spout.
3. Check all connections to make sure the underplaten tubing is securely
in place.
4. Set the waste tank back in place.
5. Correct the pump volumes, see Chapter 8.5.2.
8.6 Replacing the line fuses
You will need:
• 2 new line fuses, 250V, 1.0A, 5mm x 20mm, Type T
• 1 small, blade-type screwdriver
To replace the line fuses, proceed as follows:
Unplug system 1. On top of the system, at the right side, press the power switch.
The
green power light in the power switch goes off, and the fan and the
conveyor spirals stop.