8 Troubleshooting
8
2. At the rear of the system, set the main switch to position O, see
Figure 8-6.
3. Unplug the power cord from the power module.
Remove used line fuses 4. On the side near to the main switch, pry open the line fuse compartment
using a small, blade-type screwdriver.
5. Carefully remove the fuse holder.
6. Remove the used line fuses.
1 Socket
2 Main switch
3 Line fuse compartment
4 Power module
Figure 8-6 Removing the line fuses, rear view
1 2 3
4
WARNING
Electric shock by contact with live parts
Death or serious injury can result.
Unplug the power cord before opening the system.
NOTICE
Damage to the slide stainer due to inadequate line fuses
Slide stainer can break down.
Only use Siemens spare parts.