Hematek
®
3000 System – Instruction Manual – Version 1.3168
2. At the rear of the system, set the main switch to position O, see
Figure 8-6.
3. Unplug the power cord from the power module.
Remove used line fuses 4. On the side near to the main switch, pry open the line fuse compartment
using a small, blade-type screwdriver.
5. Carefully remove the fuse holder.
6. Remove the used line fuses.
1 Socket
2 Main switch
3 Line fuse compartment
4 Power module
Figure 8-6 Removing the line fuses, rear view
Electric shock by contact with live parts
Death or serious injury can result.
• Unplug the power cord before opening the system.
Damage to the slide stainer due to inadequate line fuses
Slide stainer can break down.
• Only use Siemens spare parts.