Hematek
®
3000 System – Instruction Manual – Version 1.3 78
• If so: Go to step 7.
• If not: Go to step 5.
Adjust sensing switch fingers 5. Support the sensing switch finger that is not upright near the center of
the wire with one hand.
6. With the other hand, bend the sensing switch finger slightly to the left or
to the right as necessary. If bending to the right, bend the sensing switch
finger no more than 7mm (1/4in) past perpendicular.
7. Lay a spare glass slide lengthwise along the rear edge of the platen and
slowly push the glass slide from right to left past each sensing switch
finger.
8. Listen for a slight click as the respective sensing switch is activated.
9. Check if the sensing switch is activated before the sensing switch finger
is lifted onto the top of the glass slide.
• If so: Go to step 12.
• If not: Go to step 10.
10. Support the sensing switch finger near the center of the wire with one
hand.
11. With the other hand, bend the sensing switch finger slightly backward no
more than 7mm (1/4in) past perpendicular.
1 Sensing switch fingers
2 Sensing switches
3 Back trough
4 Platen
Figure 8-3 Checking sensing switch fingers