8 Troubleshooting
Hematek
®
3000 System – Instruction Manual – Version 1.3 138
8
Disconnect tubes 2. Reach underneath the platen through the waste tank area, see
Figure 8-5.
3. Disconnect rinse, buffer, and stain tubes from the spouts under the
platen.
Prime pump tubing 4. Press the service button.
5. Place the free tube ends into the beaker.
6. Prime the pump tubing, see Chapter 5.2.
Process blank glass slides 7. Place the free ends of the stain and buffer tubes into separate 10mL
graduated cylinders.
1 Spouts
2 Rinse tube
3 Buffer tube
4 Stain tube
5 Service button
6 Platen
7 Waste tank area
8 Slide drawer
Figure 8-5 Waste tank area, bottom view
6
1 2 1 13
7
8 5
4