Siemens AG SPB7-250.844.01 Page 15 of 16 Mammomat
Novation
DR
Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams
X041E 2-15
LIST OF VOLTAGES AND TEST POINTS, STAND 0
1)
Standby typical value
2)
Floating voltage. Measure differentially.
Voltage Description ( typical value)
+5V Regulated DC voltage, 4.75V < +5V < 5.25V
+5V_REF DC voltage, +5V -5% < +5V_REF < +5V
+5V_DSP Regulated DC voltage, 4.75V < +5V_DSP < 5.25V
+7V Regulated DC voltage, 6.60V<+7V<7.25V
+5V_MPS Regulated DC voltage, 4.75V < +5V_MPS < 5.25V
0VD Digital ground
0VA Analog ground
0V_MPS Generator interface ground.
11V AC supply voltage, 11VAC
1) 2)
11V_MPS AC supply voltage, 11VAC
1) 2)
+5V Regulated DC voltage, 4.75V < +5V < 5.25V
+5V_M Regulated DC voltage, 4.75V < +5V_M < 5.25V
0V_W Digital ground optointerface D801 - D805
0V_M Digital ground optointerface D801 - D802
+14V Unregulated DC voltage, 18V
1)
+15V Regulated DC voltage, 14.25V < +15V < 15.75V
+24V Unregulated DC voltage, 32V
1)
+24VF Unregulated DC voltage, 32V
1)
14VAC AC supply voltage, 14 VAC
1) 2)
17VAC AC supply voltage, 17 VAC
1) 2)
17VACF AC supply for lamp
1)2)
0V_L Floating ground for LAMP voltage regulation
1)2)
-V_L Unregulated supply voltage for lamp control circuit
1)2)
24VAC AC supply voltage, 24VAC
1) 2)
24VACF AC supply voltage for the fan
1)
2)
0V Circuit ground
+5V Regulated DC voltage, 4.75V < +5V < 5.25V
+5V_R Regulated DC voltage, 4.40V < +5V_R < 5.25V
+5V_SP Regulated DC voltage, 4.85V < +5V < 5.15V
8.2V_SP Regulated DC voltage, 7.8V < +8.2V < 8.6V