Siemens AG SPB7-250.844.01 Page 4 of 10 MAMMOMAT
Novation
DR
Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams
4-4 X041E
Generator description of MAMMOMAT
Novation
DR
0
Fig. 2 Back of the stand (seen from the left)
NOV00122
PXCM unit
(Brick unit)
PB0852
HV Tank
X1, X2, X3
PB0851
HV inverter
X732
X725
X726
PB0854
Tube power
Potentiometer:
Anode drive frequency
X724
X713
PB0855
AC inverter
X712
X705
Fuse F2: Inside the module
X706
PB0850
Mains input converter
PB0853
Mains input
(Filter unit)
X700 LEDs X701 S700 X707 LEDs
DC_BUS_OK H3
DC_BUS_FAIL H4
X5, X6, X7, X8
X740 mAs-jumper
X730
LED:
HV_ON
Potentiometers:
I_OUT
HV_READ
HV_WISH
0VA
X731
X720
Potentiometers:
Bias Mo
Bias W
Testpoints:
0_VA
FIL_CURR_READ
MA_READ
X721
X722
X723
LED:
AC_ON
X710
X711
LEDs:
28V_prim
5V_RS232
-15VA
+15VA
28V_sec
5VD
GENERATOR: PLACEMENT OF MODULES,
CONNECTORS, LEDs AND POTENTIOMETERS
0