EasyManua.ls Logo

Siemens MAMMOMAT NovationDR - Viewing Information About the Detector

Siemens MAMMOMAT NovationDR
64 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
123
4
5
ABCDEFG
Siemens AG SPB7-250.844.01 Page 1 of 1 Mammomat
Novation
DR
Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams
X041E 5-6
X721
5
6
15
16
1
2
Fil_curr_act_p
MA_overcurr_p
MA_act_p
TP
0VA
TP
MA_READ
TP
FIL_CURR_READ
K1 K2
Sel_Mo_anode_p*Sel_large_focus_p*
= active low
*
Logic
circuit
Fil_curr_set
MA_set
MA_act
X720
5
6
39
40
42
X721
41
Sel_Mo_anode_p*
Sel_large_focus_p*
MA_sens_ret
MA_sense
11
12
38
MA_set_p
MA_set_n
37
MA_regul_on_p
10
9
Fil_curr_set_p
36
35
X721
Fil_curr_on
+15VA
385VDC_ret
0V_prim
+15V_prim
UCC3804
N201
385VDC
V204
T201
L201 L202 L203
+
+
Current limit
Voltage limit
7
4
6
5
2
3
1
26
25
X721
GND
Large_focus_Mo
Large_focus_W
Small_focus_Mo
Small_focus_W
BL
Generator
door
BN
WH
GY
BU
YE
PK
Bias_voltage
X726
h2W
h3
h1W
h2M
h1M
G
Bias_volt_detected_p
det.
logic
0V_sec
28V_sec
+
PWM
IC
Bias_W
R111 R109
on
off
on
off
Bias_Mo
Inv
Sel_Mo_anode_p
V_Bias
V_Bias_ret
Fil_curr_act
Filament converter
Filament focusing converter
GENERATOR - TUBE POWER SUPPLY:
PRINCIPLE DIAGRAM
TUBE FILAMENT AND FOCUS SELECTION