EasyManua.ls Logo

Siemens MAMMOMAT NovationDR - Electronic Readout

Siemens MAMMOMAT NovationDR
64 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
123
4
5
ABCDEFG
X041E
Siemens AG SPB7-250.844.01 Page 6 of 17 Mammomat Novation
DR
Medical Solutions Rev. 03 08.04 CS SD 24 Wiring Diagrams
D801 TESTPOINTS AND LED’s
X812
X813
X816
PROM
CPU
J1
V1
V2
V3
V4
AR
KVA
COLL_ST
GRID_SP
COMP_SP
SPEED
FILT_P
GRID_0
TXD
RXD
0VA1
7V
5V_REF
5V_DSP
X814
X811
F1
0VD1
5V
MPS
X815
F2
V7
V6
V5
S1
RESET SWITCH
CPU BOARD INTERNAL, GENERATOR INTERFACE 0
~
=
6
5
4
1
3
7
8
2
X811
1
2
X757
6
5
4
1
3
7
8
2
D750
MASTER
D801
(
*
KVA)
(VH)
(AR)
0V_MPS
MPS
GND
COMMUNICATION
T801
BUBN
11V_MPS
X886A
56
F2
1AT
GENERATOR INTERFACE
FLOATING, GROUNDED BY GENERATOR
V26
0V_MPS
0V_MPS
+5V_MPS
5V_MPS
TXD
RXD
+5V_MPS
0V_MPS
MPS
0V_MPS
0VD1
0VD
CPU &
PLD
FUNC.
0VA
*
AR
*
KVA
J1
X112MHz
STATUSLED’s
V1
V2
V3
V4
DATA AND ADDRESS
PROM RAM EEPROM
128k x 8 32k x 8 8k x 8
*
RESET
RESET
0VD
S1
*
RESET SW
*
RESET_HW
+5V
*
RESET
COMP
<4.6V
STATUS LED’s DURING STARTUP:
PROM/RAM TEST V1, V4 LIT.
V2, V3 LIT IF PROM CHECKSUM ERROR.
V3, V4 LIT IF RAM ERROR.
ALL TESTS PASSED, V1-V4 ARE DARK.
V60 ~ 1HZ WHEN PLD IS PROGRAMMED
X815
0V_MPS
5V_MPS
V26
FOCUS_P
ROT_POS
COLL_P2
COLL_P1
0VD2
X886
56
F802
1AT
I9
3-6
V60
ROM
ROM
PLD
V60
STAND_REQ_HV_OFF
GEN_PREP_RDY
GRID_RUNNING

Other manuals for Siemens MAMMOMAT NovationDR

Related product manuals