EasyManuals Logo

Siemens Maxum II Service Manual

Siemens Maxum II
58 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #12 background imageLoading...
Page #12 background image
Note
When removing materials from the analyzer, all items must be placed on a clean, non-abrasive
surface. Use a clean lint-free cloth.
Figure 2-7 8-Cell TCD Block Exploded View
2.2.2 Procedure to Replace Thermistor Beads
Within this procedure, the numbers in parenthesis denote parts referenced in the list contained
in the Exploded Diagram earlier in this section. Refer back to the figure for locations.
1. Shut off flows and shut down power to the analyzer. Allow the detector to cool down.
2. Remove the 4 screws and lock washers (3) that secure the cover to the detector block (4)
and remove the cover.
3. Making note of the connection location for each wire, remove the wiring (8) that is connected
to the Thermistor/Filament Board (12) that is to be removed.
Thermal Conductivity Detector
2.2 Replacing TCD Components
Maxum Edition II Detectors
12 Service Manual, May 2018, A5E42019847001

Other manuals for Siemens Maxum II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens Maxum II and is the answer not in the manual?

Siemens Maxum II Specifications

General IconGeneral
CategoryGas Chromatograph
Measurement PrincipleGas Chromatography
Communication ProtocolsModbus, Ethernet, Profibus
Operating Temperature-20°C to 55°C
Power Supply24 VDC or 100-240 VAC
Mountingrack mount
ApplicationsRefinery, chemical, petrochemical, natural gas
DetectorsThermal Conductivity Detector (TCD), Flame Ionization Detector (FID)
Sample InletsMultiple
DimensionsVaries by configuration (e.g., 19" rack mount)

Related product manuals