Parameter Value
Selectivity Components with electron affinity to carbon 10
3
to 1
Temperature Range 70 - 110°C
(higher temperature ranges on request; system
configuration dependent)
Table 5-3 Pulse Discharge Detector, Photoionization Mode
Parameter Value
Target Components Photon sensitive components
Measurement Type Mass
Smallest Measuring Range (at 60°C oven temperature) 0 to 2 ppm ±5%
Typical Measuring Range (at 60°C oven temperature) 0 to 5 ppm ±5% FS
Dynamic Range 10
3
at fixed attenuation
Linear Range 10
3
at fixed attenuation
Selectivity Photon sensitive components
(example: benzene to hexane 10
3
to 1)
Temperature Range 70 - 110°C
(higher temperature ranges on request; system
configuration dependent)
Valco Pulse Discharge Detector
5.1 PDD Specifications
Maxum Edition II Detectors
58 Service Manual, May 2018, A5E42019847001

Other manuals for Siemens Maxum II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens Maxum II and is the answer not in the manual?

Siemens Maxum II Specifications

General IconGeneral
CategoryGas Chromatograph
Measurement PrincipleGas Chromatography
Communication ProtocolsModbus, Ethernet, Profibus
Operating Temperature-20°C to 55°C
Power Supply24 VDC or 100-240 VAC
Mountingrack mount
ApplicationsRefinery, chemical, petrochemical, natural gas
DetectorsThermal Conductivity Detector (TCD), Flame Ionization Detector (FID)
Sample InletsMultiple
DimensionsVaries by configuration (e.g., 19" rack mount)