2.3 Movements :
Check arm counterbalancing at all positions.
When the single tank does not remain at the required position, the spring tension
should be adjusted for complete balancing.
Refer Troubleshooting instructions for adjustments of spring tension.
Presence of High Tension
Observe Safety Precautions
Use Radiation Protection
2.4 Exposure
Switch the generator ON.
Set the exposure parameters as 60kV, 10mAs.
Pull the recoilable Exposure Release Switch and release an exposure. The yellow
LED on the top panel will light up for duration equal to exposure time.
Audible indication by a muffled sound from single tank as well as from the buzzer will
confirm the working of the inverter and the exposure.
The LED’s on D915 glow in the following sequence.
S27 – STEP 1(Half press): V22
During exposure : V23
Error indication : V24
In case of an error the corresponding code will be displayed on the seven-segment
display. Refer the service instructions to rectify the error.