Service Manual
Installation Instructions
Siemens Ltd. Med Division India Version 4.0
Page 12 of 12
3 Handing Over Multimobil 10
Important:
Hand over the machine after taking sample exposures:
1. Chest : 60KV,12.5 mAs ( +/- 1,2 points)
2. Skull : 70KV,25mAs(+/- 2 points)
3. Extremities : 42KV,6.4 mAs (+/- 2 points)
4. Abdomen : 70KV,25mAs (+/-3 points)
After handing over the Generator, kindly fill the Delivery Confirmation Card and the
feedback report accompanying the Operating Manual and return it to the factory.
In case of any of the Following conditions, Please refer the SERVICE
INSTRUCTIONS:
Unit is in the “Sleep “ state i.e. with unit ON/OFF there is no display on the Top
panel.
Unit is working in the standby state, but proper exposure is not obtained, error is
displayed on the Top panel.
Even in the standby state, there is continuous display of error.
For ensuring the proper operation of the unit, Please refer the OPERATING
INSTRUCTIONS for the user.