5 - 1
Siemens-Elema AB Register 3 SPB7-230.033.10 Page 1 of 26 Opdima?
Solna, Sweden Rev. 03 08.02 SPS-UD Installation
Installation 5
General 5
Before starting the installation, a backup of the installation parameters of the
MAMMOMAT has to be carried out.
1. Connect a PC to the MAMMOMAT and start the service program.
2. Select Main menu => Backup => Copy installation area to disk => All.
This will store the installation parameters on the hard disk of the PC.
3. Select Main menu => Backup => Copy installation area to floppy => All.
This will store the installation parameters on the service diskette.
4. Switch off the MAMMOMAT.
Sub-assemblies, overview 5
The components shall be placed according to the figure below.
The workstation shall be placed at least 1.5 m from the object
table of the stand.
X
Y
Z
m
m
m
m
m
m
X
Y
Z
X
Y
Z
Z
X
Y
Biopsy
controller
Biopsy unit
CCD camera
Generator
Workstation
Stand with
X-ray unit
Workstation table
Fig. 1 MAMMOMAT 3000 Nova with digital biopsy and spot imaging system Opdima
?