EasyManua.ls Logo

Siemens Servo Ventilator 300 - Page 102

Siemens Servo Ventilator 300
131 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3
100 Siemens-Elema AB E382 E380E 061 01 03 01
Disassembling and assembling Servo Ventilator 300/300A
Pneumatic interconnection board
PC 1607
Separate the pneumatic section from the
power section as previously described.
Remove the cover plate as previously
described.
If an inspiratory valve – Optional (1) is
mounted, loosen the screw (2) and
remove the valve dummy.
Remove:
PC 1615 (3).
PC 1622 (4).
PC 1611 – Exp. (5).
PC 1611 – Insp. (6).
If the pressure in a pressure transducer
is raised abnormally, the transducer can
be damaged. When handling the two
pressure amplifier boards PC 1611,
never raise the pressure in the transdu-
cers by covering the hole or squeezing
the silicon rubber connection piece.
Disconnect:
Interconnection cable connectors (7 – 8).
Power supply cable connectors (9 – 10).
Safety valve connectors (11 – 12).
Expiratory valve connectors (13 – 14).
Remove the nine nuts (15).
Remove the two screws (16).
Carefully lift off PC 1607 (17).
3
4
5
6
1
2
10
13 14
17
16
11
9
7
8
15
15
12
300-A45X