EasyManua.ls Logo

Siemens Servo Ventilator 300 - Analog I;O terminal N81

Siemens Servo Ventilator 300
131 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
124 Siemens-Elema AB E382 E380E 061 01 03 01
Diagrams Servo Ventilator 300/300A
8
Analog I/O terminal N81
62 pole D-sub connector (N81). Can be used
for connection of monitoring/recording
equipment. Only Siemens connection cable
must be used.
The illustration of the connector show
outside view.
If a signal has three different scale factor
values, e. g. 3/18/45 V/l/s, the different
values refers to selected front panel setting
Adult/Pediatric/Neonate.
Input control signals can be connected at
N81 to control some ventilator functions
(e.g. when using Bi-Phasic Ventilation).
These input signals are routed via two
sockets, J2 and J3, on PC 1587.
300-A17X
21
62
42
1
43
22
N81
1 Chassis GND
2 V- limited; -15 V, max 200 mA, Output
3 0 V (A), Input/Output reference
4 V+ limited; +15 V, max 200 mA, Output
5–
6* Ext press contr lev set in, 0 V – +5 V, Input
7* Ext CMV freq set in, 0 V – +5 V, Input
8* Ext SIMV freq set in, 0 V – +5 V, Input
9* Ext insp time set in, 0 V – +5 V, Input
10 Neb flow buff 3 18 45, 3/18/45 V/l/s, Output
11 Air flow buff 3 18 45, 3/18/45 V/l/s, Output
12 Insp flow buff 3 3 18, 3/3/18 V/l/s, Output
13 Exp flow buff 3 3 18, 3/3/18 V/l/s, Output
14 Insp press buff 50 mV, 50 mV/cm H
2
O, Output
15 Exp press buff 50 mV, 50 mV/cm H
2
O, Output
16 Internal battery voltage lim, Output
17 Insp time buff, 0V ; 5 V, Output
18 Range sel D0 buff, 0V ; 5 V, Output
19 Clock buff 2 H, 0V ; 5 V, Output
20 Patient trig buff H, 0V ; 5 V, Output
21 Serial data to slave buff H, 0V ; 5 V, Output
22 GND, Input/Output reference
23 +5V limited; max 500 mA, Output
24 0V (24), Input/Output reference
25 +24V limited; max 270 mA, Output
26
27* Ext press supp lev set in, 0 V – +5 V, Input
28* Ext insp rise time set in, 0 V – +5 V, Input
29* Ext preset min vol set in, 0 V – +5 V, Input
30* Ext pause time set in, 0 V – +5 V, Input
31* Ext CPAP flow set in, 0 V – +5 V, Input
To enable the control functions, jumpers
must be mounted in the sockets. The figure
below, showing sockets J2 and J3, indicates
which connectors that need to be short-
circuited with a jumper to enable a control
signal. (Signal No. as stated in the pin
configuration list below.)
Note – If jumpers are mounted and a
voltage is connected to any of the pin
number marked * in the table below, the
function of the ventilator can be influenced.
32 Airway flow buff 3 3 18, 3/3/18 V/l/s, Output
33 Barometric press buff 4.0, 4.0 V/Bar, Output
34 ET CO
2
conc buff 1.0, 1.0 V/%CO
2
, Output
35 O
2
flow buff 3 18 45, 3/18/45 V/l/s, Output
36 Insp flow patient buff 3 3 18, 3/3/18 V/l/s, Output
37 CO
2
conc buff 1.0, 1.0 V/%CO
2
, Output
38 Serial data to master test H, 0V ; 5 V, Output
39 Range sel D1 buff, 0V ; 5 V, Output
40 Power up reset buff L, 0V ; 5 V, Output
41 Exp time buff, 0V ; 5 V, Output
42 Alarm buff L, 0V ; 5 V, Output
43
44
45
46
47* Ext O
2
conc set in, 0 V – +5 V, Input
48* Ext nebulizer time set in, 0 V – +5 V, Input
49* Ext trigg sens set in, 0 V – +5 V, Input
50* Ext PEEP lev set in, 0 V – +5 V, Input
51 Exp flow patient buff 3 3 18, 3/3/18 V/l/s, Output
52 O
2
conc buff 90 mV, 90 mV/%O
2
, Output
53 Insp flow patient buff 7.2 7.2 43.2,
7.2 /7.2/43.2 V/l/s, Output
54 Exp flow patient buff 7.2 7.2 43.2,
7.2 /7.2/43.2 V/l/s, Output
55 Insp press buff 10.7 mV, 10,7 mV/cm H
2
O, Output
56
57
58
59 Serial data to slave test H, 0 V ; 5 V, Input
60 Receive address test L, 0 V ; 5 V, Input
61 Control enable test H, 0 V ; 5 V, Input
62 Receive address buff L, 0V ; 5 V, Output
300-L27X
J3
1234567 8
9
10
111213141516
1234567 8
9
10
111213141516
J2
Signal
No:
29 8 30 9 31
47 48 49 50 27 6 28 7

Table of Contents

Related product manuals