3
102 Siemens-Elema AB E382 E380E 061 01 03 01
Disassembling and assembling Servo Ventilator 300/300A
Removing the inter-
connection cable
Patient unit connection
Remove the pneumatic section casing as
previously described.
Disconnect the two interconnection cable
connectors (1).
Remove the two screws (2).
Slide out the cable clamp (3) from the
patient unit.
Mounting the inter-
connection cable
The procedure for assembling the inter-
connection cable to the ventilator is the
reverse of the removing procedure
previously performed.
Control unit connection
Remove the two screws (1) and lift off
the cover (2).
Disconnect the two interconnection cable
connectors (3).
Remove the two screws (4).
Slide out the cable clamp (5) from the
control unit.
4
5
1
2
3
300-A48X
1
3
2
300-A47X