What to do for Siemens Servo Ventilator 300 error code µP SCM RESTART?
- DDr. Gilbert CampbellSep 12, 2025
To resolve the 'µP SCM SEE OPERATING MANUAL/ RESTART' error on your Siemens Industrial Equipment, replace PC 1587, the COM-PROM, or PC 1618 1.

What to do for Siemens Servo Ventilator 300 error code µP SCM RESTART?
To resolve the 'µP SCM SEE OPERATING MANUAL/ RESTART' error on your Siemens Industrial Equipment, replace PC 1587, the COM-PROM, or PC 1618 1.
What to do if Siemens Industrial Equipment shows 'CHECK TUBINGS'?
If you receive a 'CHECK TUBINGS' error on your Siemens Industrial Equipment, replace the expiratory pressure bacteria filter and the expiratory pressure transducer.
What to do if Siemens Industrial Equipment shows error code PF?
If your Siemens Industrial Equipment displays the technical error code 'PF SEE OPERATING MANUAL', the suggested solution is to replace PC 1618 or PC 1608.
What to do if Siemens Industrial Equipment shows error code Ba?
If your Siemens Industrial Equipment shows the technical error code 'Ba SEE OPERATING MANUAL', replace PC 1608.
How to fix Siemens Industrial Equipment error code PFT RESTART?
To resolve the 'PFT RESTART' technical error code on your Siemens Industrial Equipment, check or replace the MON-PROM. Note that the alarm activates if MON-PROM V6.0 or higher is mounted on PC 1608D or lower. You may need to replace PC 1608 or PC 1618 1.
What to do if Siemens Servo Ventilator 300 shows error code PoO?
If your Siemens Industrial Equipment displays the technical error code 'PoO SEE OPERATING MANUAL', replace the O concentration potentiometer.
What to do if Siemens Servo Ventilator 300 Industrial Equipment shows error code SwR?
If your Siemens Industrial Equipment displays the technical error code 'SwR SEE OPERATING MANUAL', replace the Patient range selector.
What to do for Siemens Servo Ventilator 300 Industrial Equipment error code µP R&T RESTART?
To address the 'µP R&T SEE OPERATING MANUAL/ RESTART' error on your Siemens Industrial Equipment, replace PC 1588 on PC 1605. You may also need to replace REF-PROM, PC 1605, or PC 1618 1.
What to do for Siemens Industrial Equipment error code µP Pan RESTART?
To address the 'µP Pan SEE OPERATING MANUAL/ RESTART' error on Siemens Industrial Equipment, replace PC 1588 on PC 1614. You might also need to replace the PAN-PROM, PC 1614, or PC 1618 1.
How to fix Siemens Servo Ventilator 300 error code ROM RESTART?
To resolve the 'ROM RESTART' technical error code on your Siemens Industrial Equipment, replace PC 1588 on PC 1608. Alternatively, replace PC 1608 or PC 1618 1.
| Manufacturer | Siemens |
|---|---|
| Model | Servo Ventilator 300 |
| FiO₂ | 21 - 100 % |
| Ventilation Modes | Volume Control, Pressure Control, SIMV, Pressure Support, CPAP |
| PEEP | 0 - 20 cmH₂O |
| Alarms | High/Low pressure, Apnea, Low minute volume, Power failure |
| Power Supply | 100-240 VAC, 50/60 Hz |
| Data Interfaces | RS-232 |
Overview of ventilator's main units and fundamental operational principles.
Detailed explanation of gas flow paths and feedback control system mechanisms.
Summary of regulation, monitoring, interface, and power supply functions.
Comprehensive description of the panel and control sections, including interfaces and monitoring.
Detailed breakdown of pneumatic section components and power section functions.
Essential safety precautions and general preparation steps before disassembly.
Procedures for handling PC boards, PROMs, and major unit sections.
Steps for selecting language, address code, and voltage settings.
Procedures for replacing fuses, batteries, and resetting memory.
Guide to identifying and resolving technical errors based on alarm messages.
Details on PROM, PC board, and operating manual version changes.
Alphabetical listing of topics with page references for easy navigation.
Visual representation of electrical connections and functional block relationships.