7ML19985JB02 SITRANS Probe LU (PROFIBUS PA) – INSTRUCTION MANUAL Page 19
mmmmm
Startup
Quick Setup
Activating SITRANS Probe LU
Power up the instrument. SITRANS Probe LU starts in RUN mode, and the LCD displays
the output of AIFB1.
Setting the PROFIBUS address via the handheld programmer
1
1. Press Mode to activate PROGRAM
mode and open Menu level 1.
2. Press Right ARROW twice to navigate
to PROFIBUS address.
3. Press Right ARROW again to open Edit
mode: the PROGRAM icon will flash.
4. Key in a new value and press Right ARROW
to accept it.
The LCD displays the new value; PROGRAM
icon disappears, and the last menu digit
flashes to indicate Navigation mode.
5. Press Mode to return to RUN mode.
Note: Keep infrared devices such as laptops, cell phones, and PDAs, away from
SITRANS Probe LU to prevent inadvertent operation.
Notes:
• Local programming must be enabled, to allow changes. (In PDM, go to Identification
> Device > Local Operation Enable and select the desired setting.)
• CLEAR can be used to clear the field.
• Press Right ARROW to open Edit mode: the PROGRAM icon flashes.
• Press Left ARROW to cancel Edit mode: the Menu number flashes (the
PROGRAM icon is not visible).
1.
Default address is 126.
C
Menu level: last
digit flashes in
Navigation mode.
Program icon:
flashes in Edit
mode.
Right-most digit
flashes in
Navigation
mode.