The LCD monitor/video display are highly sensitive to mechanical
damage.
Cleaning the camera lens
Carefully clean the camera lens with a lint-free cloth and lens
cleaner.
Cleaning the data carrier
1 Clean contaminated data carriers with a clean cloth (cotton or
microfiber).
2 Follow the manufacturer's notes when cleaning CD/DVD data
carriers.
Cleaning the plugs and connectors
1 Dampen a soft cloth with water or a diluted household cleaner
solution. Do not use organic solvents such as alcohol or acetone.
2 Carefully wipe the plugs and connectors with the cloth. Do not
touch the contacts.
Cleaning the patient table and the straps
1 Clean the patient table using a liquid household agent.
2 Wash the straps of the patient table at a temperature of
60°C.
3 Use commercial disinfecting materials. However, do not use
solutions with alcohol or acetone.
Cleaning and disinfecting the receptors
1 Do not use cleaners or disinfectants containing alcohol or ether.
2 Do not use hard or sharp objects (e.g. knives or tweezers) to
remove residue.
3 Clean the receptors with a commercially available cleaner. Follow
manufacturer’s instructions.
6.1.2
6.1.3
6.1.4
6.1.5
6.1.6
6 Maintenance
118
MR System | Operator Manual
Print No. M6-04401.621.02.02.02