CAUTION
Squeeze bulb is defective!
Risk of injury to patient because emergencies cannot be
communicated
Check the functionality of the squeeze bulb daily.
Coolant accidents
During a quench, a person might become unconscious due to severe
shortness of breath:
1 Remove unconscious persons immediately from the examination
room.
2 Start adequate first aid measures and contact a physician
immediately.
Direct contact with subzero liquids, gases, and surfaces (e.g. pipes)
may lead to frostbite. The eyes and mucous membranes are
especially vulnerable.
CAUTION
Improper handling of liquid helium!
Skin damage caused by frostbite
Do not rub frostbitten skin areas.
1 Remove clothing carefully from the locations involved.
2 Rinse frostbitten skin with lukewarm water.
3 Cover frostbitten skin with sterile bandages.
4 Do not apply powder or creams.
5 Contact a physician immediately.
2.4.3
First aid in case of shortness of
breath
First aid in case of frostbite
Safety 2
MAGNETOM Skyra
fit
| syngo MR E11
45
Print No. M6-04401.621.02.02.02