4 Use a dampened cloth for cleaning.
Do not submerge the receptors in cleaning liquid.
5 Disinfect the receptors with a commercially available disinfecting
agent. Follow the manufacturer’s instructions.
Disinfecting system components
Disinfecting sprays damage electronic components. For this
reason, components may only be cleaned with a semi-dry cloth.
WARNING
Flammable cleaning or disinfection agents may cause fire or
explosion!
Injury to patient
Disinfect the system components with commercial
disinfecting materials. However, do not use solutions with
alcohol or acetone. Follow manufacturer’s instructions.
Siemens does not test alcohol-containing, phenol-alkaline and
alkaline-based disinfectants for harmful effects on surfaces.
Avoid disinfecting system components with these materials.
Care and cleaning of floors
Do not use the following cleaning or care products:
Sprays
Silicon-based cleaning or care products
Cleaning or care products with substances that release ammonia
Cleaning or care products that destroy the anti-static properties of
the floor covering
6.1.7
6.1.8
Maintenance 6
MAGNETOM Skyra
fit
| syngo MR E11
119
Print No. M6-04401.621.02.02.02