Ensure to prevent potential current loops as shown in the red
labeled illustration.
Ensure the patient is positioned with proper distance (
5 mm) to
magnet tunnel as well as proper distance between parts of the
body as shown in the green labeled illustration.
To lower the effects of gradient fields or RF fields, keep a sufficient
distance from the RF coils and the magnet tunnel (gradient coils),
and reduce the time of exposure during measurements.
Safety 2
MAGNETOM Skyra
fit
| syngo MR E11
21
Print No. M6-04401.621.02.02.02