CAUTION
Local variation in the sensitivity of local coils!
Continuous fluctuations in MR image brightness
Whenever possible, use a local coil with transmit
characteristics that are more suitable for the field of view
desired.
Use the normalization filter.
CAUTION
Static and/or stationary brightness errors on the LCD monitor
due to aging!
Incorrect diagnosis
Scroll through the images to ensure that the MR image does
not show differences in brightness, spots, or cloudiness and
check bright objects for afterglow.
Keep your eyes always in central monitor position with a
vertical view angle towards the screen surface for best image
quality.
CAUTION
Inhomogeneous RF field!
Asymmetry of contrast in the MR image
Whenever possible, use a local coil with transmit
characteristics that are more suitable for the desired field of
view.
Variations in brightness
Variations in signal and
contrast
Safety 2
MAGNETOM Skyra
fit
| syngo MR E11
35
Print No. M6-04401.621.02.02.02