After the Magnet Stop switch has been pressed, an alarm is triggered
at the alarm box. The WARNING LED will light up and an alarm signal
will sound.
As a rule, Siemens Service must be called following a quench.
The magnet must only be put back into operation by Siemens
Service personnel.
WARNING
Formation of droplets due to condensation during quenching!
Personal injury
Risk of fire
◆ Do not touch the exhaust line.
◆ Do not stand under the exhaust line.
◆ Avoid open flames and do not smoke.
WARNING
System indicates Magnet Stop error!
Hazardous conditions because the magnet cannot be
quenched in case of emergency
◆ Immediately remove the patient from the magnet.
◆ Restrict the access to the examination room.
◆ Notify Siemens Service.
Safety 2
MAGNETOM Skyra
fit
| syngo MR E11
43