EasyManua.ls Logo

SIGLENT TECHNOLOGIES SDM3045X - Capacitance Characteristic; Temperature Characteristics

Default Icon
32 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SDM3045X Service Manual 5
Capacitance Characteristic
Accuracy±% of Reading +count
[1]
Function
Range
[2]
Max
Testing
Current
Resolution
Accuracy
one year
23 ± 5
Capacitance
2 nF
10 μA
0.001 nF
3+10
20 nF
10 μA
0.01 nF
1+10
200 nF
100 μA
0.1 nF
1+ 9
2 μF
100 μA
0.001 μF
1+10
20 μF
1 mA
0.01 μF
1+10
200 μF
1 mA
0.1 μF
1+ 9
10000 μF
1 mA
1 μF
2+50
Remarks:
[1] Specifications are for 0.5 Hour warm-up and REF operation. Using of non-film
capacitor may generate additional errors.
[2] Specifications are for from 1% to 110% on 2 nF range and ranges from 10% to 110%
on other ranges.
Temperature Characteristics
Accuracy ±% of Reading
[1]
Function
Probe
Type
Probe
Model
Working
Temperature
Range
Accuracy
one year
23 ± 5
Temperature
coefficient
018
2850
Temperature
RTD
[2]
α=0.00385
-200~660
0.16
0.09
TC
[3]
B
20~1820
0.76
0.14
E
-270~1000
0.5
0.02
J
-210~1200
0.5
0.02
K
-270~1370
0.5
0.03
N
-270~1300
0.5
0.04
R
-50~1760
0.5
0.09
S
-50~1760
0.6
0.11
T
-270~400
0.5
0.03
Remarks:
[1] Specifications are for 0.5 Hour warm-up, not include probe error.
[2] Specifications are for 4-wire measure or 2-wire measure under REF operation.
[3] Built-in cold terminal compensation for thermocouple, accuracy is ±2.

Related product manuals