Input
Error from Nominal
(1 years)
Irms
Frequency
Range
20mA
1 kHz
20 mA
±0.12 mA
10 kHz
±0.54mA
2mA
1 kHz
200 mA
±0.41 mA
200mA
1 kHz
±1.2 mA
10 kHz
±5.4 mA
20mA
1 kHz
2 A
±6.1 mA
2A
1 kHz
±14 mA
10 kHz
±54 mA
200mA
1 kHz
10 A
±26 mA
10A
1 kHz
±65 mA
10 kHz
±270 mA
Capacitance Verification
Input
Error from Nominal
(1 year)
Capacitance
Range
2 nF
2 nF
±0.07 nF
20 nF
20 nF
±0.3 nF
200 nF
200 nF
±2.9 nF
2 μF
2 μF
±30 nF
20 μF
20 μF
±300 nF
200 μF
200 μF
±2.9 μF
10000 μF
10000 μF
±250 μF