EasyManua.ls Logo

Sigma AUTOMATICS SA300 - Calibration Menu (Force Re-Learn)

Sigma AUTOMATICS SA300
34 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SA300 Installation Instructions
REV.2023-10-01
25 / 32
6.3.5. Access Control Mode
Press the MENU button on the control Board to enter the Programming Menu.
Install the door operator for required application based on installation instructions.
Adjust parameters for basic setup. (As per 6.3.3)
Follow the Access Control wiring diagram to connect all the devices to proper terminals.
See Section 5.5 for wiring details.
Signal from access control panel needs to be dry (0 Volts) to connect device and signal from access control
panel to proper terminals.
Once the devices are wired and door operator is programmed for basic setup, access the settings and
change the CONTROL MODE parameter from 1 (Regular) to 3 (Access Control).
Test to ensure proper functioning.
Once the door is closed the display will show “DOOR CLOSED [Access Denied].
6.4. Programing
Install the operator based on the following instructions and use the required tools.
Ensure the door is fully closed and no obstacles are in the doorway.
Turn Power ON. Screen will display 2 options as in image below.
Press ACT. Door will run 2 cycles. First cycle the operator learns door position. Second cycle the operator learns
door speeds.
Once proper learn is complete, screen will display DOOR CLOSED.
DOOR CLOSED
[Access Denied]