SIKA TD 32500 Operating Manual
5.2.1.4 Select physical total flow unit
• The message „SET TOTAL -UNIT – NO“ appears.
• Switch with the ▲ key to „YES“ and confirm with the
key.
• Select with the ▲ key one of the following units:
gallons per (Great Britain)
• Confirm the currently displayed unit by pressing the
key.
• Continue to configure the temperature unit, return to the menu start, or
quit the configuration menu.
5.2.1.5 Select physical temperature unit
This setting is only possible with instruments which are equipped with a
temperature measuring input.
• The message „SET TEMP-UNIT – NO“ appears.
• Switch with the ▲ key to „YES“ and confirm with the
key.
• Select with the ▲ key one of the following units:
degrees
• Confirm the currently displayed unit by pressing the
key.
• Continue to configure the pulse rate, return to the menu start, or quit the
configuration menu.
5.2.1.6 Programming pulse rate
N
ote: The pulse rate of the installed flow sensors was already set in
the factory before delivery.
• The message „SET PULSE RATE – NO“ appears.
• Switch with the ▲ key to „YES“ and confirm with the
key.
• The message „VALUE OK ?“ along with the last set pulse rate appears.
• Switch with the ▲ key to „CHANGE VALUE ?“ and confirm with the
key.
• A cursor appears. Set with the ▲ key each digit of the new pulse rate of
the connected flow sensor and confirm with the
key.
• Continue to configure the gate time, return to the menu start, or quit the
configuration menu.