SIKA TD 32500 Operating Manual
5.2.3.2 Select start value of reading range for bargraph
• The message „SET RANGE LO-VAL – NO“ appears.
• Switch with the ▲ key to „YES“ and confirm with the
key.
• The message appears „VALUE OK ?“ together with the last set low
value of the measuring range.
• Switch with the ▲ key to „CHANGE VALUE ?“ and confirm with the
key.
• A cursor appears. Set with the ▲ key the respective digits of the new
low value of the measuring range and confirm it with the
key.
• Continue to configure the end value of reading range, return to the menu
s
tart, or quit the configuration menu.
5.2.3.3 Select end value of reading range for bargraph
• The message „SET RANGE HI-VAL – NO“ appears.
• Switch with the ▲ key to „YES“ and confirm with the
key.
• The message appears „VALUE OK ?“ together with the last set end
v
alue of the measuring range.
• Switch with the ▲ key to „CHANGE VALUE ?“ and confirm with the
key.
• A cursor appears. Set with the ▲ key the respective digits of the new
end value of the reading range of the bargraph and confirm it with the
key.
• Quit the configuration menu, return to the menu start or continue to
c
onfigure the successive parameters.
5.2.4 Programming alarms
You can only carry out this configuration if TD 32500 is equipped with alarm
outputs.
• Switch over to configuration menu.
• Press the
key until the message „SET ALARM – NO“ appears.
• Switch with the ▲ key to „YES“ and confirm with the
key.
5
.2.4.1 Programming operating mode 'alarm output':
With this programming you define whether alarm 1 or 2 shall serve as
closed or open circuit current outputs.
If the alarm output 1 (2) is set as open-circuit current alarm, it will be
switched to the voltage supply in the case of an alarm, thus switching-on a
connected consumer, e.g. a relay. If no alarm is present, the alarm
output 1 (2) is separated from the voltage supply and the connected
consumer is switched-off.