APPENDIX 3
ERROR CODE SUMMARY
Error Code Explanation Recommended Action
Err.0 Sample weight too low Use large sample (greater than 115g)
Err.1 Sample cell not correctly Remount the sample cell firmly
mounted on to the interface
Err.2 Capacitance reading is Check hardware. If necessary,
too high (over 100) scale down capacitance by
changing byte 28
Err.3 Capacitance reading is Check hardware. If necessary,
too high at low moisture scale down capacitance by
changing byte 28
Err.4 High moisture reading at Check hardware. If necessary,
low capacitance reading this can be rectified by shifting
the calibration
Err.5 Capacitance reading is Check hardware.
too low for the calibration Check calibration curve
curve (i.e. %MC would be
negative)
Err.6 Communication error Check cable and start again
in RS232
batt Battery low indicator Replace the batteries