meter for blood glucose testing. Call customer service at
+86-731-89935581/+86-731-89935582 or contact local dealers for help.
4. Eject the test strip
Attention: Discard used test strip and control solution according to local regulations.
Keep control solution away from children and pets.
Questionable Results:
Blood Glucose Testing
Prepare meter, strip, lancing device and lancet before test.
22 23
Probable Cause
Operation mistake
Control solution not being shaken well
Retest according to correct steps
Shake control solution well and retest with a new test strip.
Use the first drop of control solution Retest with a new test strip
Expired or contaminated control solution Change a new vial control solution and retest.
Expired or contaminated test strip Change a new vial test strip and retest.
Temperature is too high or too low Equilibrate meter, test strip and control solution in room
temperature for 30 minutes and retest.
Meter malfunction Contact customer service or local dealers
Action
STEP 1: Wash your hands or use an alcohol swab to
clean the fingertip. MAKE SURE YOUR HAND IS
DRY BEFORE TESTING.
STEP 2: Unscrew the lancing device cover.
STEP 3: Insert a lancet into the lancing device until
it comes to a full stop.