8 9
If test results greater than 600 mg/dL or 33.3mmol/L, the meter will display “HI”. Retest immediately with a
new test strip. If your reading is still “HI”, contact healthcare professional immediately.
If test results low than 20 mg/dL or 1.1mmol/L, the meter will display “Lo”. Retest immediately with a new test
strip. If your reading is still “Lo”, contact healthcare professional immediately.
If you are experiencing symptoms that are not consistent with your blood glucose test results and you have
followed all instructions described in the safe-Accu user manual, call your healthcare professional.
A red blood cell count (hematocrit) that is either very high (above 60%) or very low (below 30%) can cause
false results.
Small molecular reducing materials vitamin C(ascorbic acid), content of vitamin C in blood exceeds
5 mg/dL can cause false results.
Content of serum uric acid exceed 10 mg/dL(0.55mmol/L) can cause false results.
High blood fat (content of cholesterol exceed 300 mg/dL(16.6mmol/L) can cause false results.
Note: Do not change your treatment only based on test result and light indications. Please consult healthcare
professional for your treatment.
Learning Your New System
The complete Safe - Accu Monitoring System
Glycemic control
Time of Day
Preprandial plasma glucose
(before a meal)
Postprandial plasma glucose
(1-2 hours after starting a meal)
70-100 mg/dL
<140mg/dl
ADA suggested reading