Memory
Your Safe-Accu Meter can store 200 most recent test results with date and time in its
memory.When the memory is full,the oldest result is dropped and the newest is added.it
also provides you with 7-,14- and 28- day averages of you blood glucose test results.
Recall the test results in memory with these easy steps.
Recalling the Latest 200 blood glucose test results
Start with the meter off,Press and hold button “M” 3 seconds to turn on the meter.The
most recent blood glucose test result appears on the display.Press and release button “D”
and the next most recent test result will appear.Keep pressing and release button “D”,the
meter will recall up to your last 200 (if there is so many results) in order from the most
recent to the oldest,then the28-,14- and7- day averages.Press and release button “C”, the
meter will recall the blood glucose test results in totally different order.Press and hold
button “M” for 3 seconds to turn off the meter.
Attention: Your may recall the test results only with no test strips inserted in the meter.
Delete test results in memory:
Start with the meter off,press and release button “M” to turn on the meter.The most
recent blood glucose test result appears on the dispaly.To delete test results in memory,-
press button “M” and button “C” together,“- - -”will appear on the dispaly when all test
results are deleted.
30 31
Information for Healthcare Professional Use
• Only trained healthcare professional be allowed to use Safe-Accu blood glucose
monitoring system on multiple patients.
• Only trained healthcare professional be allowed to collect venous blood samples.
• Any patient who has infectious diseases must use his/her own meter.
• If the same meter be used on multiple patients, there is risk of infection between both
patients and healthcare professional.
• Water or other disinfection solution which remains on skin may dilute blood and cause
inaccurate test results.
• Please dispose used lancet, test strips and expired control solution as medical waste.
Perform test:
1. Wear disposable gloves.
2. Healthcare professional use an alcohol swab to clean lance site. Then be sure dry stick
site before testing.
3. Use qualified lancing device and choose suitable penetrate depth for patients.
4. Using lancing device to lance.
5. Hold the tip of the test strip to the blood sample until the test strip reaction zone is full
and the meter begin countdown.
6. Wash hands thoroughly with soap and water after handling the meter, lancing device or
test strips.
7. Cleaning and disinfecting meter and lancing device before use on next patient.