Comparing Meter with Laboratory Results
When comparing results between Safe-Accu blood glucose monitoring system and a
laboratory system, Safe-Accu blood glucose monitoring system blood tests should be
performed within 30 minutes of a laboratory test.
Cleaning and Disinfecting
• Clean and disinfect meter immediately after getting any blood on the meter or if meter is
dirty.
• If the meter is being used on second person, the meter and lancing device should be
clean and disinfected first. Do not clean the meter during the test.
To clean the meter:
1. Wash hands thoroughly with soap and water.
2. Make sure meter is off and a test strip is not inserted. Use 75% alcohol to rub the entire
outside of the meter. Make sure no liquids enter the test strip port or other opening in the
meter.
3. Let meter air dry thoroughly before using to test.
4. Wash hands thoroughly again after handing meter.
5. Verify the meter is working properly by performing control solution test. If your meter
can’t work properly, please contact customer service at +86-731-89935581/
+86-731-89935582 or contact local dealers.
32 33
To disinfect the meter:
Please according to clinic institution disinfect method to disinfect the meter.
Power Saving
• If there is no blood applied to the test strip within 3 minutes, the meter will automatically
turn off.
After test, the test results will be displayed on the screen with test strip in strip port, If
there no operation on the meter within 3 minutes, the meter will automatically turn off.
If there is no other operation on the meter within 1 minute, such as enter into test results
review mode, the meter will automatically turn off.
Performance Characteristics
Accuracy: 95% test result of Safe-Accu meet the requirement below:
Concentration Range
<5.5 mmol/L (100 mg/dL)
≥5.5 mmol/L (100 mg/dL)
Within±0.83 mmol/L(15 mg/dL)
Within ± 15%
Bias%