Page 5
CAUTION
Use of incompatible cleaning agents
will cause damage to the fixture. Avoid
the use of cleaning solutions which
contain high concentrations of alcohol,
ethelene glycol, phenol, iodophors, or
glutaraldehyde based disinfectancts.
Staining, pitting, discoloration and
diffuser cracking may occur if these are
used.
WARNING
Be sure the power is turned "OFF" and
the bulbs have cooled before changing.
WARNING
DO NOT attempt to remove a bulb by
pulling on the glass surface or end cap.
This may cause the bulb to break off in
your hand.
CAUTION
Halogen bulbs are sensitive to body
oils. DO NOT handle the glass surface
of the bulbs as body oils from your
fingers can create a "hot spot" and may
cause the bulb to burn out prematurely.
This can best be avoided by using the
plastic wrapper that the bulb is
packaged in or a clean cloth wrapped
around the base of the bulb when
installing.
NOTE
To extend the life of the bulb reflector
surface, it should not be included in
normal cleaning. Is should be cleaned
only if absolutely necessary. Clean
gently with a clean, damp, soft cloth and
a mild soap solution. NO abrasives.
WARNING
Apply BLUE LOC-TITE to any
replacement or loose vertical support
tube attaching screws.
NOTE
•Recheck the mounting plate to make
sure it is absolutely level. In most
cases, releveling the mounting plate will
solve any drifting problems.
•If the lighthead sticks or is difficult to
move, before making any adjustments,
lubricate the BOM needle bearings with
a Teflon type spray lubricant such as
TRI-FLOW (available from Richardson-
Vicks, Inc., Memphis, TN) or equivalent.
See figure 4-10. After spraying, rotate
the lighthead several times in both
directions and recheck for proper
tension.
NOTE
The internal circuitry used in the Stellar
system requires the use of a true RMS
type digital voltmeter to accurately
set the bulb voltage. Premature bulb
failure will result from incorrect voltage.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Skytron ST19 Series and is the answer not in the manual?

Skytron ST19 Series Specifications

General IconGeneral
BrandSkytron
ModelST19 Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals