Guidance and Manufacturer’s Declaration
– Electromagnetic Emissions
Harmonic emissions IEC
61000-3-2
Voltage uctuations/icker
emissions IEC 61000-3-3
Emissions test Compliance
Group 1
Class B
Class A
Complies
RF emissions CISPR 11
RF emissions CISPR 11
Warning:
Portable RF communications equipment (including peripherals such as
antenna cables and external antennas) should be used no closer than 12
in (30 cm) to any part of the equipment, including cables specied by
the manufacturer. Otherwise, degradation of the performance of this
equipment could result.
29
Guidance and Manufacturer’s Declaration Electromagnetic Immunity
Surge IEC 61000-4-5
Power frequency magnetic eld
IEC 61000-4-8
Conduced RF IEC 61000-4-6
Radiated RF IEC 61000-4-3
NOTE: U
T is the a.c. mains voltage prior to application of the test level.
Voltage dips, short interruptions
and voltage variations on power
supply input lines IEC
61000-4-11
0% U
T; 0.5 cycle at 0°, 45°, 90°,
135°, 180°, 225°, 270°, and 315°
0% UT; 1 cycle
70% UT; 25/30 cycle
Single phase: at 0°
0% UT; 250/300 cycle
0% U
T; 0.5 cycle at 0°, 45°, 90°,
135°, 180°, 225°, 270°, and 315°
0% UT; 1 cycle
70% UT; 25/30 cycle
Single phase: at 0°
0% UT; 250/300 cycle
Immunity Test IEC 60601-1-2 Test L evel Compliance Level
±8 kV contact
±2 kV, ±4kV, ±8 kV, ±15 kV air
±2 kV for power supply lines
± 0.5 kV, ± 1 kV line(s) to line(s)
±8 kV contact
±2 kV, ±4kV, ±8 kV, ±15 kV air
±2 kV for power supply lines
± 0.5 kV, ± 1 kV line(s) to line(s)
30 A/m
50Hz/60Hz
3 V r.m.s.
150 kHz to 80 MHz
6 V RMS in the ISM and amateur
bands between 0.15 MHz and
80 MHz
10 V/m
80 MHz - 2.7 GHz
80 % AM at 1 kHz
10 V/m
80 MHz - 2.7 GHz
80 % AM at 1 kHz
3 V r.m.s.
150 kHz to 80 MHz
6 V RMS in the ISM and amateur
bands between 0.15 MHz and
80 MHz
30 A/m
50Hz/60Hz
Electrostatic discharge(ESD) IEC
61000-4-2
Electrical fast transient/burst
IEC 61000-4-4