30
Frequency Requirement
TX POWER
Frequency Band Transmit (MHz)
Band 2 1850 - 1910
1710 -1755
699 - 716
777 - 787
Receive (MHz)
1930 - 1990
2110 - 2155
729 - 746
746 - 756
Band 4
Band 12
Band 13
Frequency Band Max Power
Band 2/4/12/13
21 dBm +1.7/-3 dB
Min Power
< -39 dBm
**Pulse
Modulation:
217Hz
**Pulse
Modulation:
217Hz
**Pulse
Modulation:
217Hz
2W 28 V/m 28 V/m
0.2W 9 V/m 9 V/m
2W 28 V/m 28 V/m
Radiated RF
IEC
61000-4-3
Note* - As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because
while it does not represent actual modulation, it would be worst case.
Note** - The carrier shall be modulated using a 50 % duty cycle square wave signal.
1720 MHz
1845 MHz
1970 MHz
2450 MHz
5240 MHz
5500 MHz
5785 MHz
**Pulse
Modulation:
217Hz
2W 28 V/m 28 V/m
Guidance and Manufacturer’s Declaration-IMMUNITY to proximity elds from RF wireless
communications equipment
IEC60601 Test Level
Test Frequency Modulation
**Pulse
Modulation:
217Hz
*FM+ 5 kHz
deviation: 1kHz
sine
**Pulse
Modulation:
18Hz
1.8W
2W
0.2W 9 V/m 9 V/m
27 V/m
28 V/m
27 V/m
28 V/m
Immunity Test
Radiated RF
IEC
61000-4-3
Maximum
Power
Immunity Test
Level
Compliance
Level
385 MHz
450 MHz
710 MHz
745 MHz
780 MHz
810 MHz
870 MHz
930 MHz