6. Do not use the power supply or power cord if they are damaged,
wires are frayed or exposed. Contact your healthcare provider for a
replacement power supply and cord.
7. The RENASYS GO device is only to be used with Smith & Nephew
authorized components. Use of any other products have not been
proven safe and effective with the RENASYS GO device.
8. Check the canister viewing window regularly for signs of bleeding.
9. Canisters should be changed at least once a week or in the event that
the canister contents reach maximum volume indication (300ml or
750ml fill line).
10. RENASYS GO contains systems intended to prevent charging of
the battery at high temperatures. When charging at high ambient
temperatures (approximately 86°F/30°C), the device may suspend
charging, therefore full charge may not be achieved. In a worst-case
scenario 50% charge will still be possible (approximately 11 hours
of therapy). In this scenario ‘Charging’ will stop being displayed on
the display screen and ‘Battery Full’ on the display screen will not
be reached. In this case, the device can be used as normal while
connected to mains power, but to fully charge the battery it will be
necessary to reduce the temperature of the device, for example
by removing the carry bag and/or relocating to a lower ambient
temperature.
In the event that the device is only partially charged due to high
temperatures, battery depletion warnings (low battery, critical battery,
etc.) will continue to be operable as normal even when operating from
battery power.