37
Carry bag (available for 300ml canister)
Your device may be used with a carry bag. To place the device in the bag:
1. Open the zipper on both sides of the bag and place on a flat surface
with viewing flaps on top.
2. Place the device on top of the open section of the bag.
3. Pull the front of the bag over the top of the device and feed both parts
of one zipper together.
4. Close the zipper on the other side of the bag.
*smith&nephew
RENASYS™GO
Negative Pressure Wound
Therapy
5. Once fitted, ensure the canister tubing can move freely.
• Excess tubing can be coiled and placed in the pouch on the back
of the bag.
• Flaps on the top of the bag are for the privacy of the user. On
the rear of the bag there is a belt loop should you wish to wear
RENASYS™ GO around your waist.
RENASYS™ GO
Negative Pressure Wound
Belt Loop
Tubing
Pouch
Note: Device and canister should remain in an upright position to
maximize canister volume and ensure full alarm functionality.