25
EN
Alarm/indicator
Battery full
yrettaB lluF
m001 ***gHm
Device Status
The battery is fully charged and has up to 20 hours
therapy time remaining.
Indicator light/
audio
Battery status top indicator: Solid green
Impact to
therapy
Device is operating normally.
Steps to remedy
No action required
Precaution
Full charge and the Battery Full display may not be
possible when charging at high ambient temperatures
(approximately 86°F/30°C). It is not necessary to
continue charging for longer than 3 hours.
Alarm/indicator
Battery operating on internal battery
Device Status
Device is not plugged into an electrical outlet and battery
has up to 10 hours therapy time remaining.
Indicator light/
audio
Battery status top indicator: Flash green
Impact to
therapy
Device is operating normally.
Steps to remedy
No action required

Other manuals for smith&nephew RENASYS GO