EasyManua.ls Logo

smith&nephew RENASYS GO - Recharge and Critical Battery Alarms

smith&nephew RENASYS GO
42 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
27
EN
Alarm/indicator
y
RAHCER! EG
woL.V! rettaB
Device Status
Device is not plugged into an electrical outlet and battery
is very low.
Indicator light/
audio
Status/alarm indicator: Solid yellow
Battery status bottom indicator: Solid yellow
The audible alarm will sound.
The audible alarm cannot be paused.
Impact to
therapy
Device will continue to operate normally for 1 hour and
will stop operating once the battery is fully discharged.
Steps to remedy
Plug the device into an electrical outlet.
Therapy will continue to be delivered while the device
battery is charging.
Alarm/indicator
YR
RAHCER! EG
WOL.XE! ETTAB
Device Status
Device is not plugged into an electrical outlet and battery
has only 2 minutes of therapy time remaining.
Indicator light/
audio
Status/alarm indicator: Solid yellow
Battery status bottom indicator: Solid yellow
The audible alarm will sound.
The audible alarm cannot be paused.
Impact to
therapy
The battery has only 2 minutes of therapy time
remaining. Device must be plugged into electrical outlet
to continue therapy. If it is not plugged into an electrical
outlet it will stop operating once the battery is fully
discharged.
Steps to remedy
Plug the device into an electrical outlet.
Therapy will continue to be delivered while the device
battery is charging.