28
Settings Settings Menu
5.2.10 Thermal Camera Calibration
As the temperature of the thermal sensor influences the measured values, the imager performs
automatic calibrations during operation to calibrate the thermal imager for the current temperature
of the sensor. Selecting Thermal Camera Calibration allows you to manually perform a
calibration at anytime during operation.
Automatic Calibration
The Thermal Imager uses an automatic calibration feature to ensure consistent and accurate
temperature measurements are provided.
Additional Information:
A screen message (e.g. calibrating) is displayed during the automatic calibration process.
An automatic calibration will occur when the thermal imager is turned on during “warm-up”.
An automatic calibration will occur approximately every 5 minutes and once at every 3°C
temperature change of the thermal sensor.
An audible "click" may also be heard during calibration, and a temporary screen “freeze”
occurs.
A symbol is used to indicate calibration is about to occur (see following table):
5.2.11 Advanced
Temperature
The Temperature settings menu allows you to set the temperature unit of measure (°F or °C)
(Figure 5-15).
Use the Up / Down control buttons to make a selection from the menu, then press the Y button to
enable the selection and return to the settings menu.
Figure 5-15 Temperature Settings Menu
Example
Description
At one minute (or less) before a calibration occurs, a “~” symbol is displayed in front
of the Center Region Temperature (Average) to indicate calibration is about to occur.