GENERAL DESCRIPTION OF EQUIPMENT
D0716C-EN
10
VALID STORAGE CONDITIONS
Removing an endoscope will render the cycle invalid if:
The maximum storage time has been exceeded (set in the Settings menu: maximum value 7 days)
The air supply has been interrupted during the channel blowing cycle
The blowing time for the channels has not been respected (set in the Settings menu: minimum value
60 min)
The power supply has been interrupted since inserting the endoscope in the cabinet
There has been excess humidity and/or temperature while the endoscope was present in the cabinet
(set time and alarm threshold in the Settings menu)
The door has been open for too long while the endoscope was present in the cabinet (set time and
alarm threshold in the Settings menu)
There has been a filter failure while the endoscope was present in the cabinet (pressure threshold
exceeded)
The cabinet has been emptied while the endoscope was present in the cabinet
Any one of these conditions renders the storage invalid.
Endoscopes present in the aseptic unit when an alarm is triggered are associated with this alarm.
TECHNICAL SPECIFICATIONS & OPERATING CONDITIONS
Simple cabinet
Double cabinet
Pass-through cabinet
Dimensions (L. x D. x H.)
107 x 70 x 235 cm
184 x 70 x 235 cm
110 x 80 x 217 cm
Empty weight
300 kg
450 kg
286 kg
Consumption of medical air at full
load
40 l/min
80 l/min
40 l/min
Recommended blowing time
(channels)
60 minutes
Medical air connection
Normalised
Minimum/maximum air pressure
> 2 bar and < 7 bar
Grade of compressed air
Medical grade filtered air in accordance with standards in force
Supply voltage
220/240V 50 Hz
Power supply
600 W
Electrical connection capacity
10 A (oscillating if possible)
Operating temperature
from 15°C to 30°C (59°F 86°F) (humidity 70%)
Storage temperature
from 5°C to 50°C (41°F 122°F)
Use
Indoor
Altitude
<2000m
Pollution degree
2
Relative humidity
<70%
Overvoltage category
Category II