VALID STORAGE CONDITIONS
Removing an endoscope will render the cycle invalid if:
• The maximum storage time has been exceeded (set in the Settings menu: maximum value 7 days)
• The air supply has been interrupted during the channel blowing cycle
• The blowing time for the channels has not been respected (set in the Settings menu: minimum value
60 min)
• The power supply has been interrupted since inserting the endoscope in the cabinet
• There has been excess humidity and/or temperature while the endoscope was present in the cabinet
(set time and alarm threshold in the Settings menu)
• The door has been open for too long while the endoscope was present in the cabinet (set time and
alarm threshold in the Settings menu)
• There has been a filter failure while the endoscope was present in the cabinet (pressure threshold
exceeded)
• The cabinet has been emptied while the endoscope was present in the cabinet
Any one of these conditions renders the storage invalid.
Endoscopes present in the aseptic unit when an alarm is triggered are associated with this alarm.