INSTALLATION & DETAILED SPECIFICATIONS
D0716C-EN
11
INSTALLATION & DETAILED SPECIFICATIONS
The following information is provided to ensure the correct operation of your AS300. If this information is not
observed, cycle performance may be modified, resulting in damages that will not be covered by the
manufacturer’s warranty. We recommend that you leave the AS300 switched on permanently (display lit). If
the AS300 is switched off for a long period, please contact your authorised distributor.
For its initial operation, all interior surfaces of the cabinet must be cleaned thoroughly using a surface cleaning
and disinfecting solution.
Never place any objects on top of the cabinet. In the event of an emergency, pull out the mains plug
to disconnect the AS300 from the power supply.
INSTALLING & CONFIGURING THE AS300
Installation
The AS300 must be installed by a qualified Soluscope technician and its installation must be validated by the
installation conformity statement supplied with the AS300. This document constitutes proof of installation
qualification and validation.
It is important to position the storage unit to ensure sufficient space for normal operation, use and
maintenance:
SINGLE CABINET
DOUBLE CABINET
The overall height is 235 cm for AS300V2 and AS300x2 versions.
The overall height is 217 cm for AS300 TR version.
Ensure minimum space of one meter in front of the storage unit to facilitate use.
The AS300 must be installed close to where the endoscopes are disinfected, in a location appropriate for use
such as an endoscope decontamination room. For technical or operational reasons, it is preferable that clean
areas are located within clean rooms. Users should manage the flow of personnel and equipment in such a way
as to prevent the risk of contamination when moving from one area to another.
The AS300 unit is portable and sits on 8 feet the height of which can be adjusted. It is not necessary to attach
the device to the wall or floor.
Your electrical installation must be connected to the main earth terminal of the building at 1 wall socket: 220-
240 V + earth 50 Hz 16A via a dedicated power supply cable for the AS300.