The storage unit is equipped at the rear with a flexible cable for connection to the filtered medical grade air
source. A normalised supply of medical grade air must be installed close to the storage unit.
Installation qualification includes:
• Verification of the environment for the functioning of the unit (air, power supply).
• Physical check including parameters and all functional and alarm tests.
• Provision of a technical manual to the medical technician.
• Training of personnel using the machine.
Configuration
After checking that the cabinet functions correctly, the qualified Soluscope technician will configure various
parameters, such as:
• the date and the time,
• The process settings (maximum storage time, fan speed, etc.)
• user names,
• the list of endoscopes which will be stored in the cabinet,
Please ensure that the technician has the support of a person who will be in charge of providing the elements
required to install the unit correctly.
DESCRIPTION AND USE OF THE TOUCH SCREEN
User levels
For safety reasons, the various cabinet functions are accessed according to secure user management (user
name + passwords). Passwords correspond to different access levels and these passwords will be provided
during installation and according to your user level.
ADMIN Administrator = Qualified Soluscope technician
MAINTENANCE Soluscope technician or a technician at your hospital who has been trained by a qualified
Soluscope person
HEAD NURSE Chief Nurse at the hospital
OPERATOR Nurse
A badge containing identification information is assigned to each operator. It is read by a bar code reader (or
RFID in option) positioned on the door of the technical compartment.
Use of the touch screen
The types of endoscope are pre-recorded in the machine. When inserting an endoscope, the operator selects
the relevant endoscope reference from a drop-down menu on the touch screen.
It is possible to remove an endoscope during a ventilation cycle without compromising the storage quality.
It is not possible to remove an endoscope during a channel blowing cycle without compromising the storage
quality (invalid cycle).