Page 34
18- If the manometer reading is between 180-190 mmHg click Continue.
Note: If the manometer reading is not between 180-190 mmHg, close the SphygmoCor XCEL
Calibration software and start the calibration procedures from Step 1. Please refer to Restarting the
Calibration Procedure Section in the Calibration Troubleshooting section.